Naina Solaris Ltd

Sunday, May 20, 2012

MICRO CONTROLLER  pH-Cond-TDS-SAL Meter NIG335

MICRO CONTROLLER pH-Cond-TDS-SAL Meter NIG335

Category: pH Meter

 Function :pH & temperature OR mV & temp. OR Conductivity & temp. OR TDS & temp. OR Salinity & temp simultaneously Display : 16 character x 2 lines Alphanumeric  BL LCD Power supply :  230 V ± 10% 50 Hz AC

pH

Range : 0 to 14 pH ,Resolution: 0.01 pH, Accuracy :  ± 0.01 pH ± 1 digit, Slope adjustment:80% to 120%  By default 100%,Temp .compensation:  0 to 100 ˚C Auto with temp probe / Manual with  keys .Auto buffer recognition : 4.00 pH, 7.00 pH,9.20 pH, Calibration : Auto & manual, 3 point & 2 point (Programmable)

Conductivity-TDS-SAL

Range:  Conductivity   0.00 uS/cm to 1000 mS in 6- Auto ranges, TDS: 0.00   ppm    to 200  ppt in 6- Auto ranges , SAL:  0.0   ppt      to  50   ppt Auto ranging, Resolution: Conductivity   0.01/.1/1 u S/ m S as per range,   TDS:  0.01/.1/1 ppm/ ppt as per range, SAL:  0.1 ppt, Accuracy : ± 0.5%of range ± 1 digit, TDS Factor :   0.50(By default) & adjustable manually, SAL Factor : 0.75(By default) & adjustable manually, Cell const. adjustment :  Digital 0.05 to 2.000  By default 1.000,Temp. compensation : 0 to 70 ˚C Auto with temp. probe/Manual with  keys, Temp. Coefficient adj. : 0.00% to 4.00% adjustable. By default 2.00%, Reference Temp. :  20 Degree C & 25 Degree C, Mode of Calibration:  By Value & By Measurement

Temperature                

Range: -20.0 ˚C to 200 ˚C Resolution :  0.1 ˚C ,Accuracy :  ± 0.1% of range ±1 digit, Calibration : two point manual, Sensors : Pt100 probe, Data entry : By six soft touch keys with audible sound , Storage :  1000 samples, Real time display :  24 Hour ,Print format :All Data  & Date Wise  (Programmable),Digital output :  Serial RS-232 ,Parallel mode with Date  ,Interface: RS-232, Centronics & computer

 Print format :All Data  & Date Wise  (Programmable),Digital output :  Serial RS-232 ,Parallel Centronics (Programmable),Error Indication : pH ERROR,ORP ERROR, CAL ERROR & T-ERROR,

Accessories :  pH  Electrode , Cond /TDS Cell, Temp. Probe,  Dust  Cover, Instruction Manual.        

 

AddThis Social Bookmark Button